2.3 Specification of LAD

Incident wavelength: 0.25 - 6.5Å
Momentum transfer: The minimum achievable value is 0.25Å-1, but this value increases as backgrounds increase.
  The maximum value at which oscillations in the structure factor may be observed depends on the degree of short range order in the sample, but values in the range 30 - 50Å-1 are typical.
Beam: S7
Moderator: 100K CH4 (see section 4.6)
Sample position: 10m from moderator (L1)
Beam Size: 40mm high, 20mm wide, reducible by apertures before the sample (see section 3.4)
Count rate: 5x103 (5°) - 7x104(150°)
Sample environment: The sample tank is smaller than the ISIS standard (see section 3.1)
Data acquisition: Vaxstation 3200 (see section 4.2)

 

Bank Average Angle 2q Total Flight Path Resolution D Q/Q Maximum d-spacing Detector Type
1 4.937° 11.047m 14% ~80Å 10 atm 3He
2 9.704° 11.033m 7% ~40Å 10 atm 3He
3 20.562° 11.040m 3.5% ~17Å Li-glass scintillator
4 35.183° 11.043m 1.8% ~11Å Li-glass scintillator
5 58.103° 11.046m 1.2% ~7Å Li-glass scintillator
6 90.131° 11.039m 0.8% 4.8Å Li-glass scintillator
7 148.164° 11.128m 0.6% 3.7Å 10 atm 3He

Table 1 LAD Detector Bank Parameters

(incident flight path L1=10.0m)